A review: flexible, stretchable multifunctional sensors and actuators for heart arrhythmia therapy

Abstract

Cardiovascular disease is a very serious disease which results in about 30% of all global mortality. Atrial fibrillation (AF) causes rapid and irregular contractions resulting in stroke and cardiac arrest. AF is caused by abnormality of the heartbeat controlling electrical signal. Catheter ablation (CA) is often used to treat and remove the abnormal electrical source from the heart but it has limitations in sensing capability and spatial coverage. To overcome the limitations of the CA, new devices for improving the spatial capability have been reported. One of the most impressive methods is wrapping the heart surface with a flexible/stretchable film with an array of high-density multifunctional micro-sensors and actuators. With this technique, the overall heart surface may be diagnosed in real time and the AF may be treated much more effectively. The data acquisition from the array of multifunctional sensors is also very important for making the new devices useful. To operate the implanted device system, a battery is mostly used and it should be avoided to replace the battery by surgery. Therefore, various energy harvesting techniques or wireless energy transfer techniques to continuously feed the power to the system are under investigation. The development of these technologies was reviewed, and the current level of technology was reviewed and summarized.

Background

Arrhythmia describes an irregular heartbeat and the arrhythmia may make heartbeat rate too fast, too slow, or with an irregular rhythm [1]. There are four main types of arrhythmia: premature (extra) beats, supraventricular arrhythmias, ventricular arrhythmias, and bradyarrhythmia. Among them, the supraventricular arrhythmias are tachycardia (fast heart rates) that start in the atria or atrioventricular (AV) node and this includes atrial fibrillation, which is the most typical type of serious arrhythmia, which causes rapid and irregular atrial contractions resulting in stroke and cardiac arrest. This atrial fibrillation is caused by abnormality of the electrical signals that control the heartbeat in normal conditions [2]. By the nineteenth century, physicians had little knowledge of coronary artery disease, valvular heart disease, cardiac arrhythmias, and cardiomyopathy, although the deaths from cardiovascular disease, including arrhythmia, accounted for about 30% of all global mortality [3, 4]. In the twentieth century, they were able to diagnose and treat heart disease using new techniques including the radiograph, electrocardiograph (ECG), and cardiac catheter, etc. [3].

With better understanding of the mechanisms of AF, the cardiac ablation (CA) treatment has been more popular for the treatment of various types of AF [5]. CA removes arrhythmogenic tissue by either RF (radiofrequency) heating or cryothermy cooling with treatment by other energy sources such as laser energy, which is under clinical trials [5,6,7,8,9,10,11,12,13,14,15]. CA procedures mostly rely on a point ablation and, while it is good for treating simple arrhythmias, it is not suitable for treating complex arrhythmias that occur continuously at multiple sites due to limited sensing functionalities and single point ablation source [16]. Also, since arrhythmias can occur in all component structures and 3D regions of the heart, it is quite challenging to diagnose and treat the precise anatomic locations, especially when you want to cover a large area [6, 9]. In order to overcome this limited sensing functionality or limited spatial coverage, the catheter was designed to be inflated like a balloon as shown in Fig. 1. An advanced balloon catheter design with multiple sensors on the outside surface was reported for enabling the catheter to obtain temperature, flow characteristics, tactile, optical, and electrophysiological data from the tissue-balloon interface [17,18,19,20,21,22,23,24,25,26,27,28]. While the CA is considered as one of the best methods for diagnosing and treating AF, there is a higher risk of procedural complications due to the complexity of technique and the location of ablation sites [5]. It is reported that approximately 50,000 AF ablation procedures are being performed every year in the United States and a major complication rate of approximately 5% have been reported after CA treatment in USA [5, 29,30,31]. In order to reduce the major complications associated with the CA while increasing the sensing functionalities and spatial coverage, various novel devices that can be attached on the outer wall of the heart have been reported in order to replace the catheter for the diagnosis and treatment of the CA. One of the most impressive devices is a flexible and stretchable membrane-type device which has multifunctional sensors and actuators that can provide electrical stimulation, enabling real-time mapping of many cardiac characteristics.

Fig. 1
figure 1

(Reproduced with permission from Ref. [16], Copyright © 2011, Rights Managed by Nature Publishing Group)

Multifunctional inflatable balloon catheters. a Image of inflatable catheter (130% inflation) with contraction and (upper right) inflated state. b An enlarged image of an inflated balloon catheter (green dotted area in a)

This paper briefly reviews a recent trend in the development of flexible, stretchable multifunctional sensors and actuators for heart arrhythmia therapy that can overcome various limitations of the traditional single-point source CA technique. This new technology becomes possible by achieving improvement in the following fields—realization of flexible and stretchable substrate, successful multifunction sensors and actuators on the flexible and stretchable substrate, energy harvesting method to avoid battery replacement surgery, and information and communication technology (ICT) for exchanging the data between the sensors and other systems. Each of these subjects will be addressed with some of the recent development reports.

Next generation devices for heart arrhythmia therapy

The next generation device was aimed to diagnose and treat the complex heart arrhythmias that occur continuously at multiple sites. This complex arrhythmia cannot be treated with a single-point source catheter, even though it is an approved and one of the most widely accepted methods for treating a single point cardiac ablation due to its limited sensing functionality or limited spatial coverage. Therefore, in over to overcome these limitations various new technologies have been developed and reported. This paper introduces a part of those technologies to inform the readers what have been developed recently in the heart arrhythmia therapy field. First, in order to improve the spatial coverage of the sensors, the sensors should be fabricated on a flexible and stretchable substrate which can then be tightly and conformally attached to the curvy organ surface, in this case the heart. By fabricating small sized, high density sensors on this flexible and stretchable substrate, the device multifunctionality for collecting various cardiac characteristics can be achieved. In addition to them, the ICT capability should be included into the device for data communication and finally the power should be provided or generated without the surgery for replacing the battery.

Flexibility and stretchability

Implantable devices quite often either cause a damage to the organs or plays a source of infection by the tissue immune reaction. The implantable devices usually need to be attached tightly to the organs as long as there is no immune reaction from the organ. It is particularly critical to attach the sensors or actuators very tightly on the heart surface because the heart expands and contracts as it beats, resulting in detaching of any sensors from the surface. This will cause the failure of the sensors and with rigid sensor body it is almost impossible to have a good contact with the curvy heart surface. Even if the soft sensor body makes a good contact with the heart surface, it will be detached from the heart surface after some heartbeats because the movement of the heart is quite strong. Therefore, it is necessary to make the sensor body with a flexible and stretchable material so that the sensor body may be able to follow the shape change of the heart when the heart beats. Polydimethylsiloxane (PDMS) is one of the polymers, which Whitesides et al. reported to deform up to 200% when heating and cooling it as a stretchable structure shown in Fig. 2 [17, 32,33,34,35]. Therefore, when the sensors are fabricated on the PDMS body, they can be stretched and deformed following the heart shape without being detached.

Fig. 2
figure 2

(Reproduced with permission from Ref. [35], Copyright © 1998, Rights Managed by Nature Publishing Group)

a Gold was evaporated on PDMS at 110 °C and then cooled at room temperature to form a wave pattern. b A schematic diagram of depositing Au on PDMS and obtaining a wave pattern through heating and cooling

Multifunctional device

It is tendency to develop an array of sensors in the medical devices so that many physiological parameters can be examined at the same time from one location of the body. For example, the balloon catheter reported by Kim et al. is a multifunctional device that can measure various properties such as electrical properties, thermal properties, pH, temperature, and mechanical strain [36]. This device’s electrodes consist of a fractal structure, which allows high density sensing over a large area and low impedance without compromising elasticity and compliance [37]. Xu et al. [36] demonstrated a cardiac model with a 3D printer and created an elastic membrane called 3D multifunctional integumentary membranes (3D-MIM) as shown in Fig. 3. This device is able to diagnose various diseases of heart such as arrhythmia, ischemia, and heart failure spatially [36]. As shown in Fig. 4, Xu et al. [37] also reported an actuator capable of not only sensing but also electrical stimulation. This device has an array of 8 electrodes located around the heart circumference and it was able to deliver spatially and temporality programmed electrical stimulation onto the heart outside membrane [37].

Fig. 3
figure 3

(Reproduced with permission from Ref. [36], Copyright © 2014, Rights Managed by Nature Publishing Group)

3D-MIMs a 3D-MIM attached to the outer wall of the heart (white arrows indicate the positions of the sensors with different functions). Scale bars 6 mm. b An enlarged image of sensors with each function. Scale bars 500 mm

Fig. 4
figure 4

(Reproduced with permission from Ref. [37], © 2015 WILEY VCH Verlag GmbH & Co. KGaA, Weinheim)

An image attached to the outer wall of the rabbit heart with an elastic membrane containing a multifunctional sensor capable of providing electrical stimulation (the white arrows indicate the electrodes of the sensor and the fractal structure)

Generator

In recent years, the importance of these medical implant devices has increased dramatically and the field of application has also expanded rapidly. However, most of these devices rely on internal batteries and the lifetime of the devices is limited by the battery life [38,39,40]. The implantable medical device usually employs a small high-capacity battery with low-power devices in order to increase the lifetime of the medical device to avoid the surgery for replacing the battery or medical device. However, with the current technology, the lifetime of battery-operated pacemakers, for example, is reported to be about 5–15 years and the device replacement must be done, normally by surgery [38, 41]. Research is underway to overcome these drawbacks by considering a generator with an energy harvesting method instead of a battery [38]. Energy harvesting may be done by various methods and all kinds of energy harvesting principles are studies by researchers. Some people are trying to obtain energy using a fuel cell by a chemical reaction between glucose and oxygen, and many people are trying to harvest substantial amount of energy from various mechanical body movements including a knee, body movement, and even a heart wall movement from the heartbeats [38, 42,43,44,45,46,47,48]. Figure 5 shows an example of obtaining energy by attaching mass imbalance oscillation generator (MIOG) to the left ventricle wall of Swiss alpine sheep. From this experiment, the heart was beating at an average of 90 bpm and the experiment lasted for a total of 18 min and 45 s, producing 11.1 μJ per heartbeat [38]. However, attaching the energy harvesting device directly on the heart may cause impairment or even damage to the heart chronically. The possibility of causing cardiac damage, such as bleeding, has been raised because of heart’s vigorous contraction and vulnerability [49]. Figure 6 shows a flexible and implanted PVDF piezoelectric energy harvesting generator film wrapped on an ascending aorta for harvesting energy from the pulsation of the aorta to avoid the cardiac damage. Figure 6 shows before and after of PG injection in the pig’s aorta. Figure 7 shows that the schematic illustration of the piezoelectric generator (PG) wrapped around the latex tube, its cross-sectional SEM image, and the photograph of the fabricated the PG by depositing Al on both sides of the PVDF. The reported results showed that the average heart rate of pigs was 120 bpm and the aorta had a strain rate of about 10% due to heartbeat [49,50,51]. The instantaneous output of the PG was 30 nW and lasted for 700 ms and was charged to 1.0 V for 40 s for a 1 μF capacitor in the charge test [49].

Fig. 5
figure 5

(sheep Reproduced with permission from Ref. [38], © Biomedical Engineering Society 2012)

Attachment of mass imbalance oscillation generator (MIOG) to the left ventricle of Swiss alpine

Fig. 6
figure 6

(Reproduced with permission from Ref. [49], Copyright © 2015 Elsevier Ltd. All rights reserved)

a PG picture wrapped in PI tape. b PG implanted in the aorta

Fig. 7
figure 7

(Reproduced with permission from Ref. [49], Copyright © 2015 Elsevier Ltd. All rights reserved)

a A schematic diagram of a piezoelectric generator (PG) surrounding the aorta. b cross-sectional SEM image of PG. c PG structure diagram

Information and communication technology

Even if there is a sensor that can map every part of the heart spatially and a generator that can operate the sensor indefinitely, it is meaningless if it cannot send the collected information to the outside. In order to monitor heart information in real time, the collected information should be transmitted to an information processing device outside the body [52]. Hammond et al. developed a device that transmits heart pressure data wirelessly and can monitor the information in real time in 2012 [53]. The device, as shown in Figs. 8 and 9, encapsulates pressure sensors made with MEMS technology and electronic components that provide signal conditioning, power management, and radio frequency transmission into glass capsules, which are then implanted into the left ventricular apex. Data from this device is passed through a proprietary transceiver unit and a handheld antenna. The transceiver can provide power to the device through electromagnetic induction, detect and process radio frequency signals emitted by the device, and store the data in a computer via USB [53]. However, until now, the transmittable distance is short and only the data about the pressure can be collected and encoded and transmitted. Therefore, it is necessary to secure the data of the multi-function sensor to be able to process. Also, the connection to the sensor should be considered to process the data in the membrane type sensor which surrounds the whole heart.

Fig. 8
figure 8

(Reproduced with permission from Ref. [53], Copyright © 2012 by the American Society for Artificial Internal Organs)

a Proprietary transceiver unit and a handheld antenna. b Diagram implanted in apex of left ventricle. c Pressure sensors made with MEMS technology and electronic components providing signal conditioning, power management and radio frequency transmission are encapsulated in glass encapsulation devices

Fig. 9
figure 9

(Reproduced with permission from Ref. [53], Copyright © 2012 by the American Society for Artificial Internal Organs)

a Pictures of devices implanted in the left ventricular apex. b Implant site

Conclusions

Compared to a single point source CA, which is currently used to treat cardiac arrhythmia, a new implantable, flexible and stretchable membrane-type device with multifunctional sensors and actuators has been developed and reported in order to map in real-time, and large-area spatial cardiac characteristics. In order to operate such devices by attaching it on an organ in a live body for a long time, the battery replacement surgery should be avoided. To overcome this difficulty, various energy generation/harvesting techniques have been reported, including a fuel cell using a chemical reaction between glucose and oxygen, energy harvesting from various mechanical body movements including a heart wall movement from the heartbeats. One of the energy harvesting experiment results is that 11.1 μJ energy per heartbeat was harvested from the 90 bpm heartbeats for 18 min and 45 s by attaching the generator to the heart outside wall of the sheep. In addition, there must be a communication device capable of sending the data collected by the sensors to an external device enabling real-time monitoring outside the body. Transmitting an electrical signal through a human body suffers significant signal, limiting a real-time data transmission only for a short distance. Therefore, some of the developments in those important issues were reviewed and reported in this paper to address what needs to be developed further to make successful flexible, stretchable multifunctional sensors and actuators for heart arrhythmia therapy.

Abbreviations

CA:

cardiac ablation

AF:

atrial fibrillation

RF:

radiofrequency

3D-MIM:

3D multifunctional integumentary membranes

AV:

atrioventricular

ECG:

electrocardiograph

ICT:

information and communication technology

PDMS:

polydimethylsiloxane

MIOG:

mass imbalance oscillation generator

PG:

piezoelectric generator

MEMS:

microelectromechanical systems

References

  1. National Institutes of Health (2011) What is an arrhythmia? https://www.nhlbi.nih.gov/health/health-topics/topics/arr. Accessed 27 Mar 2017

  2. National Institutes of Health (2011) Types of arrhythmia. https://www.nhlbi.nih.gov/health/health-topics/topics/arr/types. Accessed 27 Mar 2017

  3. Tan SY, Kwock E (2016) Paul Dudley White (1886–1973): pioneer in modern cardiology. Singap Med J 57(4):215

    Article  Google Scholar 

  4. Mehra R (2007) Global public health problem of sudden cardiac death. J Electrocardiol 40(6):S118–S122

    Article  Google Scholar 

  5. Maan A, Shaikh AY, Mansour M, Ruskin JN, Heist EK (2011) Complications from catheter ablation of atrial fibrillation: a systematic review. Crit Pathw Cardiol 10(2):76–83

    Article  Google Scholar 

  6. Kim DH et al (2012) Electronic sensor and actuator webs for large-area complex geometry cardiac mapping and therapy. Proc Natl Acad Sci USA 109(49):19910–19915

    Article  Google Scholar 

  7. Dewire J, Calkins H (2010) State-of-the-art and emerging technologies for atrial fibrillation ablation. Nat Rev Cardiol 7(3):129–138

    Article  Google Scholar 

  8. Crandall MA, Bradley DJ, Packer DL, Asirvatham SJ (2009) “Contemporary management of atrial fibrillation: update on anticoagulation and invasive management strategies,” in Mayo Clinic Proceedings, vol. 84(7). Elsevier, pp. 643–662

  9. Calkins H et al (2007) Heart rhythm society; European heart rhythm association; European cardiac arrhythmia society; American college of cardiology; American heart association; society of thoracic surgeons. Europace 9(6):335–379

    Article  Google Scholar 

  10. Wittkampf FH, Nakagawa H (2006) RF catheter ablation: lessons on lesions. Pacing Clin Electrophysiol 29(11):1285–1297

    Article  Google Scholar 

  11. Arentz T et al (2003) Feasibility and safety of pulmonary vein isolation using a new mapping and navigation system in patients with refractory atrial fibrillation. Circulation 108(20):2484–2490

    Article  Google Scholar 

  12. Haissaguerre M et al (2002) Mapping and ablation of idiopathic ventricular fibrillation. Circulation 106(8):962–967

    Article  Google Scholar 

  13. Haïssaguerre M et al (2000) Electrophysiological end point for catheter ablation of atrial fibrillation initiated from multiple pulmonary venous foci. Circulation 101(12):1409–1417

    Article  Google Scholar 

  14. Greenspon AJ (2000) Advances in catheter ablation for the treatment of cardiac arrhythmias. IEEE Trans Microw Theory Tech 48(12):2670–2675

    Article  Google Scholar 

  15. Haissaguerre M et al (1998) Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med 339(10):659–666

    Article  Google Scholar 

  16. Kim DH et al (2011) Materials for multifunctional balloon catheters with capabilities in cardiac electrophysiological mapping and ablation therapy. Nat Mater 10(4):316–323

    Article  Google Scholar 

  17. Kim D-H, Lu N, Huang Y, Rogers JA (2012) Materials for stretchable electronics in bioinspired and biointegrated devices. MRS Bull 37(03):226–235

    Article  Google Scholar 

  18. Slepian MJ, Ghaffari R, Rogers JA (2011) Multifunctional balloon catheters of the future. Intervent Cardiol 3(4):417–419

    Article  Google Scholar 

  19. Lau M et al (2010) A theoretical and experimental analysis of radiofrequency ablation with a multielectrode, phased, duty-cycled system. Pacing Clin Electrophysiol 33(9):1089–1100

    Article  Google Scholar 

  20. Reddy VY et al (2009) Visually-guided balloon catheter ablation of atrial fibrillation. Circulation 120(1):12–20

    Article  Google Scholar 

  21. Meissner A et al (2009) First experiences for pulmonary vein isolation with the high–density mesh ablator (HDMA): a novel mesh electrode catheter for both mapping and radiofrequency delivery in a single unit. J Cardiovasc Electrophysiol 20(4):359–366

    Article  Google Scholar 

  22. Di Biase L et al (2009) Relationship between catheter forces, lesion characteristics, “popping”, and char formation: experience with robotic navigation system. J Cardiovasc Electrophysiol 20(4):436–440

    Article  Google Scholar 

  23. Chun K-RJ et al (2009) The ‘single big cryoballoon’technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre study. Eur Heart J 30(6):699–709

    Article  Google Scholar 

  24. Chun K-RJ et al (2009) Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. J Cardiovasc Electrophysiol 20(11):1203–1210

    Article  Google Scholar 

  25. Reddy VY et al (2008) Balloon catheter ablation to treat paroxysmal atrial fibrillation: what is the level of pulmonary venous isolation? Heart Rhythm 5(3):353–360

    Article  Google Scholar 

  26. Mansour M et al (2008) Initial experience with the Mesh catheter for pulmonary vein isolation in patients with paroxysmal atrial fibrillation. Heart Rhythm 5(11):1510–1516

    Article  Google Scholar 

  27. Schmidt B et al (2007) Pulmonary vein isolation by high-intensity focused ultrasound: first-in-man study with a steerable balloon catheter. Heart Rhythm 4(5):575–584

    Article  Google Scholar 

  28. Satake S et al (2003) Usefulness of a new radiofrequency thermal balloon catheter for pulmonary vein isolation. J Cardiovasc Electrophysiol 14(6):609–615

    Article  Google Scholar 

  29. Cappato R et al (2010) Up-dated worldwide survey on the methods, efficacy and safety of catheter ablation for human atrial fibrillation. Circ Arrhythmia Electrophysiol 2(1):10–17

    Google Scholar 

  30. Oral H et al (2002) Pulmonary vein isolation for paroxysmal and persistent atrial fibrillation. Circulation 105(9):1077–1081

    Article  Google Scholar 

  31. Oral H et al (2002) Clinical significance of early recurrences of atrial fibrillation after pulmonary vein isolation. J Am Coll Cardiol 40(1):100–104

    Article  MathSciNet  Google Scholar 

  32. Lacour SP, Chan D, Wagner S, Li T, Suo Z (2006) Mechanisms of reversible stretchability of thin metal films on elastomeric substrates. Appl Phys Lett 88(20):204103

    Article  Google Scholar 

  33. Lacour SP, Jones J, Wagner S, Li T, Suo Z (2005) Stretchable interconnects for elastic electronic surfaces. Proc IEEE 93(8):1459–1467

    Article  Google Scholar 

  34. Lacour SP, Jones J, Suo Z, Wagner S (2004) Design and performance of thin metal film interconnects for skin-like electronic circuits. IEEE Electron Device Lett 25(4):179–181

    Article  Google Scholar 

  35. Bowden N, Brittain S, Evans AG, Hutchinson JW, Whitesides GM (1998) Spontaneous formation of ordered structures in thin films of metals supported on an elastomeric polymer. Nature 393(6681):146–149

    Article  Google Scholar 

  36. Xu L et al (2014) 3D multifunctional integumentary membranes for spatiotemporal cardiac measurements and stimulation across the entire epicardium. Nat commun 5:3329

    Google Scholar 

  37. Xu L et al (2015) Materials and fractal designs for 3D multifunctional integumentary membranes with capabilities in cardiac electrotherapy. Adv Mater 27(10):1731–1737

    Article  Google Scholar 

  38. Zurbuchen A et al (2013) Energy harvesting from the beating heart by a mass imbalance oscillation generator. Ann Biomed Eng 41(1):131–141

    Article  Google Scholar 

  39. Paradiso JA, Starner T (2005) Energy scavenging for mobile and wireless electronics. IEEE Pervasive Comput 4(1):18–27

    Article  Google Scholar 

  40. Mateu L, Moll F (2005) Review of energy harvesting techniques and applications for microelectronics (Keynote Address). In: Microtechnologies for the New Millennium. International Society for Optics and Photonics; 2005, pp 359–373

  41. Kleemann T et al (2007) Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation 115(19):2474–2480

    Article  Google Scholar 

  42. Karami MA, Inman DJ (2012) Powering pacemakers from heartbeat vibrations using linear and nonlinear energy harvesters. Appl Phys Lett 100(4):042901

    Article  Google Scholar 

  43. Wang Z, Leonov V, Fiorini P, Van Hoof C (2009) Realization of a wearable miniaturized thermoelectric generator for human body applications. Sens Actuators A 156(1):95–102

    Article  Google Scholar 

  44. Romero E, Warrington R, Neuman M (2009) Energy scavenging sources for biomedical sensors. Physiol Meas 30(9):R35

    Article  Google Scholar 

  45. Kerzenmacher S, Ducrée J, Zengerle R, Von Stetten F (2008) Energy harvesting by implantable abiotically catalyzed glucose fuel cells. J Power Sour 182(1):1–17

    Article  Google Scholar 

  46. Platt SR, Farritor S, Garvin K, Haider H (2005) The use of piezoelectric ceramics for electric power generation within orthopedic implants. IEEE/ASME Trans Mechatron 10(4):455–461

    Article  Google Scholar 

  47. Tashiro R, Kabei N, Katayama K, Tsuboi E, Tsuchiya K (2002) Development of an electrostatic generator for a cardiac pacemaker that harnesses the ventricular wall motion. J Artif Organs 5(4):0239–0245

    Article  Google Scholar 

  48. Goto H, Sugiura T, Harada Y, Kazui T (1999) Feasibility of using the automatic generating system for quartz watches as a leadless pacemaker power source. Med Biol Eng Compu 37(3):377–380

    Article  Google Scholar 

  49. Zhang H et al (2015) A flexible and implantable piezoelectric generator harvesting energy from the pulsation of ascending aorta: in vitro and in vivo studies. Nano Energy 12:296–304

    Article  Google Scholar 

  50. Dagdeviren C et al (2014) Conformal piezoelectric energy harvesting and storage from motions of the heart, lung, and diaphragm. Proc Natl Acad Sci 111(5):1927–1932

    Article  Google Scholar 

  51. Bussy C, Boutouyrie P, Lacolley P, Challande P, Laurent S (2000) Intrinsic stiffness of the carotid arterial wall material in essential hypertensives. Hypertension 35(5):1049–1054

    Article  Google Scholar 

  52. Merchant FM, Dec GW, Singh JP (2010) Implantable sensors for heart failure. Circ Arrhythmia Electrophysiol. 3:657–667

    Article  Google Scholar 

  53. Hammond RL et al (2012) A wireless and battery-less miniature intracardiac pressure sensor: early implantation studies. Asaio J 58(1):83–87

    Article  Google Scholar 

Download references

Authors’ contributions

SJK and JJP wrote the manuscript. Both authors read and approved the final manuscript.

Authors’ information

Seung-Jo Kang received the B.S. degree in advanced materials science and engineering at Daejin University, Pocheon, Korea, in 2015. Currently, he is working toward the combined MS/PhD degree in the School of Electrical Engineering at Korea University, Seoul. His research interests include micro/nano-systems (MEMS/NEMS), electrochemical biosensors and energy harvesting using nano-structured materials.

James Jungho Pak received B.S., M.S., and Ph.D. degrees in electrical engineering respectively in 1985, 1988, and 1992 at Purdue University. He had worked in Intel Corporation in Santa Clara, CA, USA from 1992 to 1995 as a senior device physic–cist. Since 1995, he has been a professor in the School of Electrical Engineering at Korea University. His research interests include the microsystems including bio-MEMS, biosensor, applications of polymer in micro-sensors and actuators, flexible electronics and novel semiconductor devices and processing.

Acknowledgements

Not applicable

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

The data and materials are available from the corresponding author of this paper.

Funding

This work was supported by the R&D program of MOTIE/KEIT. [10054488, Development of cellular metabolic rate analyzer that can simultaneously measure multiple parameters (pH, dissolved oxygen, and metabolic heat) from less than 1000 cells].

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to James Jungho Pak.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kang, SJ., Pak, J.J. A review: flexible, stretchable multifunctional sensors and actuators for heart arrhythmia therapy. Micro and Nano Syst Lett 5, 22 (2017). https://doi.org/10.1186/s40486-017-0055-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40486-017-0055-9

Keywords